Yes sample prep for MagNet is very important. We noticed a huge decrease in IDs using serum instead of plasma. Anecdotally, even relatively small changes in centrifugation speed when the blood is collected seems to make a difference.

21.02.2025 15:50 β€” πŸ‘ 0    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

The Mag-Net data surprises me. We’ve used it for several projects now and usually get 3-4k proteins. We are also using an Astral but with only a 12 min gradient (20 min injection to injection).

21.02.2025 15:03 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

Orbi res doesn’t matter at all (or barely at all- there is slightly more overhead between an Orbi and Astral scan that when doing just Astral scans). We do a 240k res MS1 scan which takes 0.6 seconds, so that’s what we set the cycle time to.

18.02.2024 23:47 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 0    πŸ“Œ 0

Website hasn’t been updated in a while. 72 SPD run is $140 for internal customers. But thanks for the reminder to nudge IT again to get the new instrument and rates on the website.

18.02.2024 21:45 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 0    πŸ“Œ 0

No we pack our own columns and use a trap as well. Since getting the Astral we pack 15 cm 100 um ID columns with 1.8 um beads and flow at 600 nL/min. Accuracy was quite incredible when testing a 2 proteome mix at different spike-in levels. Probably getting about 5-6 DPPP from base to base.

18.02.2024 14:16 β€” πŸ‘ 0    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

@WelleKevin on twitter. I unfollowed almost everyone on there and just use Lists now which makes it much better for me. My DMs are open there though.

15.02.2024 14:08 β€” πŸ‘ 0    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

Happy to talk in more detail via DM on twitter if that would be helpful.

14.02.2024 23:50 β€” πŸ‘ 0    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

My guess is 6k proteins is pretty trivial at 100 SPD. We usually run at 72 SPD (20 minutes injection to injection- about 10 minutes of which is active gradient) and get about 8500 proteins from HeLa. We actually just got 10k proteins from a 72 SPD method with astroglia cells.

14.02.2024 23:50 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 3    πŸ“Œ 0

We routinely use DIA to analyze pull-downs. I try to avoid TMT most of the time, but regardless I think our DIA results for pulldowns have been great.

14.11.2023 00:35 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 0    πŸ“Œ 0

I just spoke to a collaborator in Denmark and he said they use World Courier for overseas shipping. Apparently they re-supply dry ice during shipment and packages arrive in 2-4 days. Maybe worth looking into.

09.11.2023 02:32 β€” πŸ‘ 3    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0
Preview
Proteome birthdating reveals age-selectivity of protein ubiquitination bioRxiv - the preprint server for biology, operated by Cold Spring Harbor Laboratory, a research and educational institution

Thrilled to share my first pre-print !! www.biorxiv.org/content/10.1...

07.11.2023 15:25 β€” πŸ‘ 17    πŸ” 3    πŸ’¬ 1    πŸ“Œ 1

Could use the Astral for this. Keep doing continuous 500k resolution scans in the Orbi while the Astral is handling all MS2 scans. That’s pretty much how the Astral runs now, but the default MS1 res is 240k.

27.10.2023 13:18 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

We’ll hook you up with one. We are doing a bunch of testing this week. If we do some runs on low-input lysates with multi-CV FAIMS we can send you a raw file to play around with.

22.10.2023 11:47 β€” πŸ‘ 0    πŸ” 0    πŸ’¬ 2    πŸ“Œ 0

Won’t really be able to use FAIMS for normal DIA runs on the Astral as switching between voltages takes about 25 ms. We’ll be using it primarily for TMT and for single-cell/low-input samples where injection times will be higher.

21.10.2023 17:55 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

And maybe TMT experts can correct me if I’m wrong, but I don’t think TMT will be able to drastically shorten gradients or get away with not fractionating. Isolation interference will always be an issue if compressing those gradients or increasing sample complexity by not fractionating.

19.10.2023 08:56 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 0    πŸ“Œ 0

We actually would do 2 hour runs on our 8 TMT fractions because we found SPS-MS3 was so slow, and to reduce co-isolation interference. With DIA runs becoming so short with next-gen instruments my instinct is that even with 16 samples DIA will use far less instrument time than a fractionated TMTpro16

19.10.2023 08:54 β€” πŸ‘ 2    πŸ” 0    πŸ’¬ 2    πŸ“Œ 0

For these low abundance peptides, don’t co-eluting ions have a disproportionate impact on getting accurate quan values? When doing a 3 proteome mix with known ratios, it was striking to me how poor the accuracy was for low abundant proteins. This was on a Lumos with SPS-MS3, but without RTS.

12.10.2023 00:08 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

Maybe we just didn’t spend enough time on optimizing TMT settings on our Lumos, but we got much better data with DIA than TMT. This was with a 3 proteome mix with known ratios. DIA identified far more proteins, and with far better quantitation. TMT samples were fractionated into 8 fractions too.

12.10.2023 00:01 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

Wow that’s great, especially if a one time charge. We get charged $200/TB/year for storage that is guaranteed to be backed up.

07.10.2023 20:29 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

How much does your University charge for space? We’d go bankrupt if we bought 2 Petabytes!

07.10.2023 17:06 β€” πŸ‘ 0    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

Make sure the freight elevators can handle the length of the crates. 7 feet long and we needed to go through multiple different elevators and hallways because of where our lab is. Also the crate for accessories is almost as big as the astral crate.

06.10.2023 22:51 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 0    πŸ“Œ 0

We actually had 3 guys delivering it. Thank god because it was perilous to get it through the hallways and the multiple elevators. Two massive crates. Took almost 3 hours total to get both outside our lab.

06.10.2023 22:48 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 0    πŸ“Œ 0

I didn’t want people to think I was posting about a new floor centrifuge! Install should happen on the 16th if all goes well.

06.10.2023 22:26 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 0    πŸ“Œ 0
Post image

Look what just arrived! Can’t wait for this to be installed in a couple of weeks! #astral #teammassspec

06.10.2023 21:27 β€” πŸ‘ 14    πŸ” 1    πŸ’¬ 3    πŸ“Œ 0

How long does it take to re-equilibrate those columns? That is 4.4 uL of column volume, and using 5 volumes for equilibration would bring it up to 22 uL. Are you re-equilibrating at 1000 bar so it only takes a couple of minutes? I assume the normal flow rate for that column is around 1.5 uL/min?

06.10.2023 13:07 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

You are not alone. It’s like wtf went wrong with the BCA/digestion/labeling where there could be 2x more protein in one sample vs another?

28.09.2023 13:56 β€” πŸ‘ 1    πŸ” 0    πŸ’¬ 1    πŸ“Œ 0

@neely.bsky.social sent me one already! Thanks for the repost.